[[CM Series]], [[TEM Optics]]
Via [[PSU CM200 Bringup Log]]
### Two ways of moving the spot/image
## Shift knobs in D mode
Before touching shift knobs:
![[Pasted image 20230322162907.png|med]]
Moving the spot with shift X-Y knobs while in D mode -> Beam shift coils changed
![[Pasted image 20230322162951.png|med]]
## DIFF ALIGN
In D mode with the spot centered, before touching DIFF ALIGN:
![[Pasted image 20230322163028.png|med]]
After moving the pattern across the screen with the multi function knobs while in DIFF ALIGN -> Image shift coils changed:
![[Pasted image 20230322163139.png|med]]