[[Compustage]] ## X Limit Adjustment ### Loosen this set screw It's a 1.3mm hex for the lock screw, and 1.5 mm for the limit screw. ![[IMG_9643.jpg|grid]]![[Pasted image 20230503165719.png|grid]] ![[IMG_9662.jpg]] ![[IMG_9673.jpg]] ### Prepare to move the limit screw This screw (1.5mm hex) is inserted or retracted to change the position of the x limit, and the set screw holds it in place. ![[Pasted image 20230503165649.png]] ![[IMG_9638.jpg]] ![[CompuStage Limit Adjustment 2023-05-03 17.02.15.png]] %%[[CompuStage Limit Adjustment 2023-05-03 17.02.15.excalidraw|🖋 Edit in Excalidraw]], and the [[CompuStage Limit Adjustment 2023-05-03 17.02.15.dark.png|dark exported image]]%% ![[CompuStage Limit Adjustment 2023-05-03 18.57.45.png]] %%[[CompuStage Limit Adjustment 2023-05-03 18.57.45.excalidraw|🖋 Edit in Excalidraw]], and the [[CompuStage Limit Adjustment 2023-05-03 18.57.45.dark.png|dark exported image]]%% ### Center the X range At low mag, move the stage to the minimum X position and then to the maximum X position. Adjust limit screw until the range of movement is centered on the view screen - Counterclockwise: -X - Clockwise: +X ![[CompuStage Limit Adjustment 2023-05-03 17.29.20.png]] %%[[CompuStage Limit Adjustment 2023-05-03 17.29.20.excalidraw|🖋 Edit in Excalidraw]], and the [[CompuStage Limit Adjustment 2023-05-03 17.29.20.dark.png|dark exported image]]%%