[Cleaning Procedure Outline for [[JSM 6300 - JEOL W SEM]]
Aim 1: Removal of dirt, grease, dust, and other contaminants by making target JSM-6400 systems accessible to cleaning.
Aim 2: Deoxidization of necessary CEC's for longer lifespan and ease with future trouble shooting/diagnostics.
Aim 3: Cable management through routing, cleaning, and labeling for work access.
Aim 4: Original JSM-6400 service record created for reference in future maintenance and to recap services provided; Aka the ERC or Engineer's Report Card.
==Pre-contact Requirements List:==
- Safety procedures and supplies
- Stabilized environment for JSM 6400
- System locations mapping outside and inside
- JEOL 6400 manual + existing obsidian notes
- Toolbox/cart
- Organized and accessible work bench
==Tools and Supplies==
(Toolbox and Mobile Cart continents)
- Shop vac + mini nozzle
- D5 Dioxide spray
- NBSKS (Nuts, Bolts, and Screws keeper)
- Microfiber cloth towels
- Tex wipes
- Mild soap and water spray mix
- Color coordinated zipties
- Cable combs
- Power Hammer Degreaser
- CERCT (card edge receptacle cleaning tool)
- Zep Fast 505
- Gloves/Eye protection
- ERC document
- Moving/Loading supply kit
==Breaking Into Steps==
1. Unload (moving procedure)
2. Staging and Stabilization
3. Assessment (aka first look)
4. Labeling/organization prep
5. Toolbox
6. Deconstruction for accessibility
7. First pass/surface cleaning
8. First pass cable management
9. Card cleaning/Chip re-seating
10. Connector electrical contacts cleaning
11. Next step assessment (if required)
12. Final pass, gen maintenance
13. Reconstruction, final touches
14. Secure (loading procedure)
15. Document, edit, and send ERC to client
16. Follow up and finalize payment