## Location of ion pumps
![[IMG_2953.png|med]]
Pumping speed: 60 L/s and 20 L/s
## Connections
SI4 = GUN INT SIP VAC READY
Current measurement points:
- SI9 = Gun
- SI10 = Int
## Vacuum levels
Ultimate pressure, after bakeout: Gun <= 3 x 10-8 Pa, Int <= 1 x 10-6 Pa
![[Screen Shot 2022-12-30 at 4.00.26 PM.jpg|sm]]
SIP HV Ready / Vacuum OK: Gun <= 7 x 10-8 Pa, Int <= 1 x 10-4 Pa
![[Screen Shot 2022-12-30 at 4.00.11 PM.jpg|sm]]
Protection (SIP Off): Gun/Int <= 5 x 10-2 Pa
Specimen chamber READY light: Gun <= 1 x 10-4 Pa
After SIP on for 2 hours: Gun = 4 x 10-8 Pa, Int = 1 x 10-5 Pa
Ion pump startup: Gun = 1 x 10-5 Pa
READY light on Bakeout Control: 1x10-4 Pa
## Adjustment points
TP1, TP2 = Reference voltages
VR2 = Meter adjustment, A
VR3 = Meter adjustment, Pa
VR4 = SIP current changeover
VR5 = PROTECTION adjustment
VR6 = READY adjustment
VR7 = Adjust meter and Pa LED
SW1 = Upper: Check HT OFF, Lower: HT ON
### Pressure adjustment
SIP CONT PB VR1 - set to 1 x 10-7A for 6400F, 2 x 10-7A for 6300F
- VR1 is on the back side of the PCBs
![[IMG_2954.png]]