<< [[Vacuum Technology MOC]]
[[Vacuum Pump Exhaust Filter Sourcing]]
[[Ulvac G100 rotary vacuum pump rebuild]]
[[Ulvac G-100 Series Rotary Vacuum Pump]]
![[_Users_isabelburgos_Documents_Obsidian_Notes_bin_Linked Files_Screen Shot 2021-07-20 at 5.31.58 PM.jpg|med]]![[_Users_isabelburgos_Documents_Obsidian_Notes_bin_Linked Files_Screen Shot 2021-07-20 at 5.33.52 PM.jpg|med]]
valve B = discharge
valve D = gas ballast
## Gas ballast
Without gas ballast, some gas vapors condense to liquid while building pressure at valve B. With gas ballast open, the pressure of the added gas allows valve B to open faster and pump out vapors without leaving condensed liquid.
- Cleans contaminated pump oil, but runs at a higher temperature and exhausts more oil mist
## Getting tubes off of vacuum barbs without risking cutting a gash in the barb
Use a pair of channel pliers to grip the tube right past the end of the barb. Each time you compress the tube it should slide off the front of the barb a little bit.
![[IMG_1307.jpg]]